ASTM E711 PDF

Find the most up-to-date version of ASTM E at Engineering ASTM E() [ Withdrawn ]. Standard Test Method for Gross Calorific Value of Refuse-Derived Fuel by the Bomb Calorimeter. Designation: E – 87 (Reapproved ) Standard Test Method for. Gross Calorific Value of Refuse-Derived Fuel by the Bomb Calorimeter1 This standard.

Author: Digis Bragor
Country: Singapore
Language: English (Spanish)
Genre: Video
Published (Last): 24 October 2008
Pages: 367
PDF File Size: 3.88 Mb
ePub File Size: 15.19 Mb
ISBN: 414-2-84647-524-7
Downloads: 17272
Price: Free* [*Free Regsitration Required]
Uploader: Nikokora

RDF products are frequently non-homogeneous. The crystals shall be pelletized before use.

ASTM E – 87 Gross Calorific Calorimeter – PDF Free Download

Examine the bomb interior and discard the test if unburned sample or sooty deposits are found. The immersed portion of the stirrer shall be coupled to the outside through a material of low heat conductivity. The water may asstm measured volumetrically if it is always measured at the same temperature. Allow 5 min for attainment of equilibrium; then record the calorimeter temperatures Note 6 at 1-min intervals for 5 min.

The maximum difference between any two test points shall not be more than 0. The usual range of weight is 0. Made to help students memorize the basics of muscles.

ASTM E (Re-approved ) – International Standards

Controlled room temperature and humidity are desirable. Record the initial temperature and fire the charge. Knowledge of this value is essential in assessing the commercial worth of the fuel and to provide the basis of contract between producer aztm user.

After the rapid-rise period about 4 to 5 minrecord temperatures at 1-min intervals on the minute until the difference between successive readings has been constant for 5 min. A Wheatstone bridge and galvanometer capable of measuring resistance to 0. This pressure must remain the same for qstm calibration and for each calorific determination. Firing Circuit – A 6 to 16V alternating or direct current is required for ignition purposes with an ammeter or pilot light in the circuit to indicate when current is flowing.

  ALGORITMI SI STRUCTURI DE DATE PDF

The buret used for the HNO3 titration shall be of such accuracy that estimations to 0. The crystals shall be pelleted before use.

Save my name, email, and site URL in my browser for next time I post a comment. Determine the energy equivalent of the calorimeter as the average of a series of ten individual runs, made over a period of not less than 3 days or more than 5 days.

This is the number that is multiplied by the corrected temperature rise in degrees and divided by the sample weight in grams to give the gross calorific value after thermochemical corrections have been applied.

Caution – The ignition circuit switch shall be of momentary double-contact type, normally open, except when held closed by the operator. If, by accident, the oxygen introduced into the bomb should exceed the specified pressure, do not proceed with the combustion. Collect the washings in a beaker and titrate the washings with standard carbonate solution. Recommended hygienic practices include use of gloves when handling RDF and washing hands before eating or smoking.

Sample Holder – Samples shall be burned in an open crucible of platinum, quartz, or acceptable base-metal alloy.

The maximum difference in correction between any two test points shall not be more than 0. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Base metal alloy crucibles are acceptable if after a few preliminary firings the weight does not change significantly between tasks. The weight of solid fuel sample and the pressure of the oxygen admitted to the bomb must not exceed the bomb manufacturer’s recommendations.

Do not record time intervals since they are not critical in the adiabatic method. The standard solution 0. Determine the correction for HNO3, and firing wire.

Remember me Forgot password? The amount of water g is usual can be most satisfactorily determined by weighing the calorimeter vessel and water together on a balance. Either isothermal or adiabatic calorimeter jackets may be used. Threads on the main closure should be checked frequently for wear. Gross Anatomy Cheatsheet Gross Anatomy.

  HUMAN SEXUALITY HOCK 3RD EDITION PDF

Detach the filling connection and exhaust the bomb in the usual manner. Leave a Reply Cancel reply Your email address will not be published. It shall have a device for stirring the water thoroughly and at a uniform rate, but with minimum heat input.

ASTM E711-87 (Re-approved 2004) – International Standards

Oxygen – free of combustible matter. It shall be constructed so that any water evaporating from the jacket will not condense on the calorimeter. These thermometers shall be tested for accuracy against a known standard preferably by the National Bureau of Standards at intervals asgm greater than 2. Allow 5 min for attainment of equilibrium; then record the calorimeter temperatures at 1min intervals for 5 min.

The laboratory sample must be air-dried and particle size reduced to pass a 0. Observations, Adiabatic Jacket Method – Assemble the calorimeter in the jacket and start the stirrer. Combustible waste fraction densified compressed into the form of pellets, sligs, cubettes, or briquettes.

This shall have a lens and holder designed so as to introduce no significant errors due to parallax.

The immersed portion of the stirrer shall be coupled to the outside through a material of low heat conductivity. Karl Gross juegos teorias.

The chairman of E A more diluted standard solution may be used for higher sensitivity. Collect the washings in asgm beaker and titrate the washings with standard carbonate solution.

Oxygen manufactured from liquid air, guaranteed to be greater than Valves, gauges, and gaskets must meet industry safety codes.